@article{Presipitasi48090, author = {Ayunda Sean and Etih Hartati and Dyah Marganingrum}, title = {Comparison between Calibration and Addition Method of Lead and Chromium Total Testing in Textile Industrial Wastewater using AAS}, journal = {Jurnal Presipitasi: Media Komunikasi dan Pengembangan Teknik Lingkungan}, volume = {19}, number = {3}, year = {2022}, keywords = {AAS, addition method; calibration method; textile industry}, abstract = { Most of the wastewater from the textile industry in Indonesia has a negative impact on the environment and society. The wastewater contains heavy metals such as lead (Pb) and total chromium (Cr-Tot) from the coloring process. Pb and Cr-Tot parameters testing usually use Atomic Absorption Spectrophotometer (AAS). The purpose of this study was to compare between the calibration and the addition method to test the metal parameters of Pb and Cr-Tot contained in wastewater of textile industry of PT. X. Calibration method is often use for heavy metal concentration test than addition method. However calibration method is not frequently used for low concentration while addition method can be used for that. The difference between two methods is in the procedure. The concentrations of Pb and Cr-Tot using calibration method were 0.058 mg/L and 0.018 mg/L, respectively. Meanwhile, the concentrations of Pb and Cr-Tot using the addition method were 1.1419 mg/L and 1.1036 mg/L, respectively. Based on this research, the addition method gives a relatively better value and fulfills the interval requirements in the AAS readings for low Pb metal concentrations. }, issn = {2550-0023}, pages = {651--658} doi = {10.14710/presipitasi.v19i3.651-658}, url = {https://ejournal.undip.ac.id/index.php/presipitasi/article/view/48090} }
Refworks Citation Data :
Most of the wastewater from the textile industry in Indonesia has a negative impact on the environment and society. The wastewater contains heavy metals such as lead (Pb) and total chromium (Cr-Tot) from the coloring process. Pb and Cr-Tot parameters testing usually use Atomic Absorption Spectrophotometer (AAS). The purpose of this study was to compare between the calibration and the addition method to test the metal parameters of Pb and Cr-Tot contained in wastewater of textile industry of PT. X. Calibration method is often use for heavy metal concentration test than addition method. However calibration method is not frequently used for low concentration while addition method can be used for that. The difference between two methods is in the procedure. The concentrations of Pb and Cr-Tot using calibration method were 0.058 mg/L and 0.018 mg/L, respectively. Meanwhile, the concentrations of Pb and Cr-Tot using the addition method were 1.1419 mg/L and 1.1036 mg/L, respectively. Based on this research, the addition method gives a relatively better value and fulfills the interval requirements in the AAS readings for low Pb metal concentrations.
Article Metrics:
Last update:
Last update: 2025-12-14 16:20:44
Copyright includes the exclusive right to publish and provide articles in all forms and media, including reprints, photographs, microfilms, and the like, as well as translations. The publication of every part of this journal, storage in databases, and distribution such as electronic, electrostatic, and mechanical copies, photocopies, records, magnetic media, etc., will be permitted only with written permission from Jurnal Presipitasi, Diponegoro University. Jurnal Presipitasi, Diponegoro University, and the editors and Editorial of the International Advisory Board make every effort to ensure that no false or misleading data, opinions, or statements are published in the journal. The contents of articles published in Jurnal Presipitasi, Diponegoro University, are the sole and exclusive responsibility of each author.
View My Stats
Jurnal Presipitasi is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.